Abstract
In this study, the analysis method of trace element in pure indium using glow discharge mass spectrometry (GD-MS) has been established by correcting the analysis result, which was obtained from non-standard quantitative method of GD-MS. The effect of instrument operating parameters, the resolution and isotopes and the pre-sputtering time on the determination results were investigated respectively. The optimized operating parameters of the instrument were obtained as follows: the discharge current was 37.5mA and discharge airflow was 475ml/min. Mass spectral interferences have been eliminated using selected resolution and isotope. The pre-sputtering time was 30 min. 4 trace elements in two pure samples were determined for 6 times. The RSD were less than 20%. The content of trace elements in indium samples was determined by GD-MS and Chemical Analysis Method of Indium which is tested by inductively coupled plasma atomic emission spectrometry, respectively. The results were consistent with each other.
Citation
APA:
(2016) Quantitative Analysis Of The Trace Elements In Purity Indium Material By Glow Discharge Mass SpectrometerMLA: Quantitative Analysis Of The Trace Elements In Purity Indium Material By Glow Discharge Mass Spectrometer. The Minerals, Metals and Materials Society, 2016.