RI 8853 - An Analyzer for Determining Helium-4 in the Parts-Per-Billion Range

The National Institute for Occupational Safety and Health (NIOSH)
Philip W. Holland
Organization:
The National Institute for Occupational Safety and Health (NIOSH)
Pages:
14
File Size:
617 KB
Publication Date:
Jan 1, 1984

Abstract

The Bureau of Mines has developed a trace helium analyzer for deter-mining helium-4 in the parts-per-billion range. The instrument is used for analyzing pure gases, laboratory-prepared standards, and atmospher-ic samples. Accuracy at the 95-pct confidence level in the 50-to 20,000-ppb range is within ±2 ppb or ±0.4 pct, whichever is greater. The minimum detectable helium-4 concentration is calculated to be 5 ppb. The analyzer consists of a modified helium leak-detector mass spec-trometer, a laboratory-built sample inlet system, and a liquid-nitrogen-cooled charcoal trap. A programmable integrator and event-control module make automatic operation possible. The portability of the analyzer allows it to be used in the field.