RI 9010 - An Improved Analyzer for Determining Helium-4 in Parts-Per-Billion Range

The National Institute for Occupational Safety and Health (NIOSH)
Charles Seitz
Organization:
The National Institute for Occupational Safety and Health (NIOSH)
Pages:
12
File Size:
552 KB
Publication Date:
Jan 1, 1986

Abstract

The Bureau of Mines has modified its parts-per-billion helium-4 analyzer to improve the analytical capabilities of the instrument. The modifications resulted in the following improvements: (1) The analytical precision of the instrument for 10 consecutive analyses was improved to ±1 ppb at the 5,000-ppb helium-4 level, (2) the helium-4 sensitivity was increased almost tenfold, and (3) the minimum detectable limit for helium-4 was improved to 0.5 ppb. The modifications made to improve the instrument included replacing the ion-source control circuit with a direct-current solid-state design, controlling the temperature of the ion source, stabilizing and reducing the pumping rate of the vacuum system, and optimizing peak integration.
Citation

APA: Charles Seitz  (1986)  RI 9010 - An Improved Analyzer for Determining Helium-4 in Parts-Per-Billion Range

MLA: Charles Seitz RI 9010 - An Improved Analyzer for Determining Helium-4 in Parts-Per-Billion Range. The National Institute for Occupational Safety and Health (NIOSH), 1986.